Risk Control Certifies Rates, Not Magnitudes: An Audit of Selective Fetal Biometry
首次提交 2026-11-04 10:41
版本更新 2026-11-04 10:41
雷达观测 2026-11-05 10:08
摘要
Prenatal ultrasound biometry must return a head circumference (HC) and refuse when the image cannot support one. Selective prediction trades coverage against error, but its finite-sample guarantees bound the rate of gross errors, not their magnitude. We audit a tool-grounded biometry pipeline on two separately pre-registered four-way splits of the 999 labelled HC18 images, retraining the segmenter three times on each. Calibrating the contour-completeness threshold by exact Clopper-Pearson risk control returns a certificate indistinguishable from a hand-set floor, yet the pipeline still emits a 271.3 mm error: a small, geometrically complete ring fitted to an image whose true HC is 324.6 mm. The accepted candidate scores 0.94 completeness, so the score cannot see the failure, and one outlier among $\approx 150$ answers leaves the certified risk bound ($α= 0.10$) intact. A second, untouched split reproduces the mechanism on a different image (233.3 mm). Constraining the admissibility region---the anatomical range in which the measurand is defined---removes the family instead: raising the lower bound from 50 to 55 mm costs 0.4 coverage points and cuts the worst-case error on both splits to 11.4 and 11.1 mm, whereas suppressing it through the completeness threshold costs 19.0 and 12.1 points. At an anatomically defined 70 mm the frozen tests give 1.91 and 1.81 mm MAE.
作者备注:10 pages, 2 figures, 3 tables
观测到的版本
相邻记录可能跳过中间版本;首次发现 v2 及以上也会标为版本更新。
v1 · 2026-11-04 10:41 · 最新观测版本
Risk Control Certifies Rates, Not Magnitudes: An Audit of Selective Fetal Biometry
Prenatal ultrasound biometry must return a head circumference (HC) and refuse when the image cannot support one. Selective prediction trades coverage against error, but its finite-sample guarantees bound the rate of gross errors, not their magnitude. We audit a tool-grounded biometry pipeline on two separately pre-registered four-way splits of the 999 labelled HC18 images, retraining the segmenter three times on each. Calibrating the contour-completeness threshold by exact Clopper-Pearson risk control returns a certificate indistinguishable from a hand-set floor, yet the pipeline still emits a 271.3 mm error: a small, geometrically complete ring fitted to an image whose true HC is 324.6 mm. The accepted candidate scores 0.94 completeness, so the score cannot see the failure, and one outlier among $\approx 150$ answers leaves the certified risk bound ($α= 0.10$) intact. A second, untouched split reproduces the mechanism on a different image (233.3 mm). Constraining the admissibility region---the anatomical range in which the measurand is defined---removes the family instead: raising the lower bound from 50 to 55 mm costs 0.4 coverage points and cuts the worst-case error on both splits to 11.4 and 11.1 mm, whereas suppressing it through the completeness threshold costs 19.0 and 12.1 points. At an anatomically defined 70 mm the frozen tests give 1.91 and 1.81 mm MAE.
查看此版本解读 →查看此版本原文 ↗