A Synchronized Tri-Modality Wireless System for Multi-Point Non-Contact Cardiac Monitoring
首次提交 2026-11-05 10:50
版本更新 2026-11-05 10:50
雷达观测 2026-11-06 10:04
摘要
Cardiac physiological monitoring is fundamental to evaluating cardiovascular health and diagnosing related diseases. Non-contact radar sensing can capture chest-wall vibrations with spatial specificity, enabling localized physiological monitoring and cardiac assessment. Progress, however, has been limited by the absence of a reliable, research-grade hardware platform for rigorous validation of radar-based methods. Prior systems typically pair radar, namely Doppler cardiography (DCG), with only electrocardiography (ECG) or other coarse references, providing no direct mechanical ground truth and thus constraining fine-grained evaluation of radar outputs. We present a tri-modality wireless system that integrates ECG, seismocardiography (SCG), and DCG under a unified acquisition and control framework for time-aligned recording and analysis. The tri-modality system combines miniaturized wearable ECG/SCG modules with a high-resolution mmWave radar module, supporting spatially resolved sensing and subsequent feature extraction. To our knowledge, this is the first integrated ECG--SCG--DCG hardware system for comprehensive cardiac monitoring; it enables multi-point thoracic acquisition, achieves precise cross-modal synchronization, and maintains specifications suitable for wearable operation. Experiments on a linear-motion platform show close agreement between DCG- and SCG-derived displacements, confirming accurate displacement measurement. Human-subject studies further demonstrate cross-modal consistency and robust extraction of feature points, establishing ECG as the electrical anchor and SCG as the mechanical ground truth for DCG. Overall, the system provides a rigorous, spatially resolved validation platform that advances non-contact cardiac sensing and supports next-generation radar-based monitoring.
作者备注:11 pages, 9 figures
观测到的版本
相邻记录可能跳过中间版本;首次发现 v2 及以上也会标为版本更新。
v1 · 2026-11-05 10:50 · 最新观测版本
A Synchronized Tri-Modality Wireless System for Multi-Point Non-Contact Cardiac Monitoring
Cardiac physiological monitoring is fundamental to evaluating cardiovascular health and diagnosing related diseases. Non-contact radar sensing can capture chest-wall vibrations with spatial specificity, enabling localized physiological monitoring and cardiac assessment. Progress, however, has been limited by the absence of a reliable, research-grade hardware platform for rigorous validation of radar-based methods. Prior systems typically pair radar, namely Doppler cardiography (DCG), with only electrocardiography (ECG) or other coarse references, providing no direct mechanical ground truth and thus constraining fine-grained evaluation of radar outputs. We present a tri-modality wireless system that integrates ECG, seismocardiography (SCG), and DCG under a unified acquisition and control framework for time-aligned recording and analysis. The tri-modality system combines miniaturized wearable ECG/SCG modules with a high-resolution mmWave radar module, supporting spatially resolved sensing and subsequent feature extraction. To our knowledge, this is the first integrated ECG--SCG--DCG hardware system for comprehensive cardiac monitoring; it enables multi-point thoracic acquisition, achieves precise cross-modal synchronization, and maintains specifications suitable for wearable operation. Experiments on a linear-motion platform show close agreement between DCG- and SCG-derived displacements, confirming accurate displacement measurement. Human-subject studies further demonstrate cross-modal consistency and robust extraction of feature points, establishing ECG as the electrical anchor and SCG as the mechanical ground truth for DCG. Overall, the system provides a rigorous, spatially resolved validation platform that advances non-contact cardiac sensing and supports next-generation radar-based monitoring.
查看此版本解读 →查看此版本原文 ↗