← 研踪 PaperTrailarXiv:2611.30928 · v1
Personalization of a Commercial Powered Knee Prosthesis Under Clinically Realistic Conditions via Reinforcement Learning: A Feasibility Study
首次提交 2026-11-24 00:17
版本更新 2026-11-24 00:17
雷达观测 2026-11-24 22:55
摘要
Personalization of powered prosthesis control to individual user needs remains a significant challenge. To address this challenge, policy iteration with constraint embedded (PICE), a reinforcement learning (RL) tuning approach, has been developed and validated. This feasibility study investigated whether the developed approach can personalize a commercial powered knee prosthesis under clinically realistic conditions. To reduce the time for personalization, tuning was initialized with a policy learned offline from one non-disabled participant walking on a treadmill and was then applied to three other non-disabled participants walking with the knee through a bypass adapter on a 5-meter walkway. All participants reached the goal within 8-18 tuning steps (1.7-4.2 minutes of walking), reducing the knee-kinematic feature error by 64-90% on average. These results support the feasibility of implementing the RL-based tuning approach in clinically realistic conditions with space and equipment constraints.
作者备注:6 pages, 5 figures
观测到的版本
相邻记录可能跳过中间版本;首次发现 v2 及以上也会标为版本更新。
v1 · 2026-11-24 00:17 · 最新观测版本
Personalization of a Commercial Powered Knee Prosthesis Under Clinically Realistic Conditions via Reinforcement Learning: A Feasibility Study
Personalization of powered prosthesis control to individual user needs remains a significant challenge. To address this challenge, policy iteration with constraint embedded (PICE), a reinforcement learning (RL) tuning approach, has been developed and validated. This feasibility study investigated whether the developed approach can personalize a commercial powered knee prosthesis under clinically realistic conditions. To reduce the time for personalization, tuning was initialized with a policy learned offline from one non-disabled participant walking on a treadmill and was then applied to three other non-disabled participants walking with the knee through a bypass adapter on a 5-meter walkway. All participants reached the goal within 8-18 tuning steps (1.7-4.2 minutes of walking), reducing the knee-kinematic feature error by 64-90% on average. These results support the feasibility of implementing the RL-based tuning approach in clinically realistic conditions with space and equipment constraints.
查看此版本解读 →查看此版本原文 ↗