跳到正文
研踪PaperTrail
← 研踪 PaperTrail

arXiv:2406.14234 · v3

Feed-forward active magnetic shielding

首次提交 2024-06-20 19:57
版本更新 2026-11-02 01:49
雷达观测 2026-11-04 23:43
arXiv ↗PDF ↗

摘要

Measurement of small magnetic fields requires shielding. For example, magnetoencephalography (MEG) measures small fields arising within the brain, in the presence of environmental fields several orders of magnitude larger. A common solution is passive shielding, but its cost, bulk and limited effectiveness motivate the alternative approach of active shielding, which comes in two flavours: feed-back and feed-forward. In feed-back shielding (the most common design), coils are driven from sensors that measure the corrected field. In feed-forward shielding (less common), coils are driven from sensors that measure the uncorrected field. The latter has the potential of providing better performance, but its implementation is hobbled by an inevitable coupling between coils and reference sensors. This paper describes an approach that employs a data-driven ``decoupling matrix" to suppress the coupling, enabling greater flexibility to achieve effective shielding across a wide frequency range. The method is evaluated by simulation, first using a simplified "flat world" in which the principle is clearly visualised, and then based on the more realistic geometry of a standard MEG system. Performance appears to be limited only by spatial sampling and numerical resolution. The method, which is data-driven and does not require geometric calculations, tight-tolerance fabrication, or manual calibration, is complementary with standard feed-back shielding methods. Active shielding is of particular interest in the context of a new generation of optically-pumped magnetometers (OPM), as the lower cost and reduced constraints of these new systems relative to SQUID-based systems provide an incentive to likewise reduce the cost and constraints of shielding.

AI 论文解读

基于摘要 · 非全文审阅

此版本尚未生成 AI 解读。解读由后台任务生成并按论文版本保存,登录后共享已有结果。

版本差异

目前只观测到一个版本,无法比较变化。之后收录新版本时,会在这里对照标题、作者、分类和摘要。

观测到的版本

相邻记录可能跳过中间版本;首次发现 v2 及以上也会标为版本更新。

v3 · 2026-11-02 01:49 · 最新观测版本

Feed-forward active magnetic shielding

Measurement of small magnetic fields requires shielding. For example, magnetoencephalography (MEG) measures small fields arising within the brain, in the presence of environmental fields several orders of magnitude larger. A common solution is passive shielding, but its cost, bulk and limited effectiveness motivate the alternative approach of active shielding, which comes in two flavours: feed-back and feed-forward. In feed-back shielding (the most common design), coils are driven from sensors that measure the corrected field. In feed-forward shielding (less common), coils are driven from sensors that measure the uncorrected field. The latter has the potential of providing better performance, but its implementation is hobbled by an inevitable coupling between coils and reference sensors. This paper describes an approach that employs a data-driven ``decoupling matrix" to suppress the coupling, enabling greater flexibility to achieve effective shielding across a wide frequency range. The method is evaluated by simulation, first using a simplified "flat world" in which the principle is clearly visualised, and then based on the more realistic geometry of a standard MEG system. Performance appears to be limited only by spatial sampling and numerical resolution. The method, which is data-driven and does not require geometric calculations, tight-tolerance fabrication, or manual calibration, is complementary with standard feed-back shielding methods. Active shielding is of particular interest in the context of a new generation of optically-pumped magnetometers (OPM), as the lower cost and reduced constraints of these new systems relative to SQUID-based systems provide an incentive to likewise reduce the cost and constraints of shielding.

查看此版本解读 →查看此版本原文 ↗